Skip to content

Stop the previous sensor monitor when a new one is adopted - #28

Merged
ps2 merged 1 commit into
mainfrom
fix/monitor-release-on-repair
Oct 31, 2026
Merged

ps2 merged 1 commit into
mainfrom
fix/monitor-release-on-repair

Conversation

@ps2

@ps2 ps2 commented Oct 30, 2026

Copy link
Copy Markdown
Contributor

Re-pairing a sensor without relaunching Loop left the previous LibreLoopSensorMonitor running:

  • adopt(monitor:) replaced self.monitor without stopping the old one.
  • delete() set monitor = nil without calling stop().
  • A stopped monitor's notification stream ends on cancellation, and it still called its disconnect handler. The handler then stopped the current monitor and scheduled another reconnect.

Each re-pair added another live watchdog (CCCD re-arms, patch status reads) and another source of spurious disconnects. Log lines and connect requests multiplied, and readings had long gaps until Loop was relaunched.

Changes:

  • stop() drops the monitor's handlers.
  • adopt stops the monitor it replaces.
  • delete stops its monitor.
  • onDisconnect ignores a monitor that is no longer current.

Builds in a next-dev workspace, and the LibreLoop tests pass. Not yet reproduced on hardware.

Re-pairing without relaunching left the old monitor running: adopt
replaced it without stopping it, and delete dropped it without stopping
it. Its silence watchdog kept re-arming CCCDs and reading patch status,
and when its stream ended it reported a disconnect that tore down the
live monitor and scheduled another reconnect, so log lines and connect
requests multiplied with each re-pair until Loop was relaunched.

stop() now drops the monitor's handlers, adopt stops the monitor it
replaces, delete stops its monitor, and a disconnect from a monitor that
is no longer current is ignored.
@ps2
ps2 merged commit 329301a into main Oct 31, 2026
@ps2

ps2 commented Oct 31, 2026

Copy link
Copy Markdown
Contributor Author

Tested in the iOS Simulator against libre3sim, with next-dev Loop. Scenario: Delete CGM, then Add CGM and pair the same sensor; do that twice, then have the sensor drop the link.

Without this PR With this PR
monitor reported disconnect; clearing and reconnecting for one drop 2 (consecutiveNoStreamCycles 1, then 2) 1
requesting connect for one drop 2 1
Monitors left over from deleted managers still running: patchStatus quiet …; issuing direct read every minute, and the reads time out on dead sessions stopped at delete: no patch-status reads in the 100+ s after the drop

The spurious "Replace sensor" in the original report is a separate issue (#29): an event-log entry with errorData 8 repeated in patch status while patchState is still 4. libre3sim reproduces it with POST /event {"errorData":8}. This PR doesn't change it.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment