This issue was raised by Claude (Anthropic's AI assistant, via Claude Code) on behalf of a Trio user. It was first reported as nightscout/Trio#1698; the Trio maintainers asked for it to be raised here because LibreLoop is a submodule they can't fix directly. Claude did the analysis from the user's exported log and screenshots, the code it links to, and two open-source Libre 3 decoders (DiaBLE and Juggluco). The findings were checked against the log and source, but not reproduced on hardware.
Summary
A FreeStyle Libre 3 Plus showed Replace Sensor / Sensor failed for 2 h 33 min (13:36 → 16:09 AEDT, 30 Oct 2026), while the sensor:
- reported
patchState4 (active) in every status packet, - kept sending readings (117
dq=goodreadings after 13:36), and - kept the host app (Trio) looping normally.
The state never cleared on its own; only pairing a new sensor cleared it. The user removed a working sensor at 5.6 of its 15 days.
Important
Root cause in short: bytes 0–6 of the Libre 3 patch-status packet are the sensor's most recent event-log entry, not its live condition. Around 13:22 the sensor logged an entry with errorData = 8. RoundWhiteDiscKit maps errorData 8 → replaceSensor without looking at the entry's age or at patchState, and LibreLoop persists sensorNeedsReplacement. Every later packet repeats the same entry, and LibreLoop only re-evaluates when the attention value changes, so the latch can't clear until the sensor logs a different event or a new sensor is paired.
Environment
| Sensor | FreeStyle Libre 3 Plus (15-day), 5.6 days into wear |
| Host app | Trio 1.0.2 (53), main @ 0b29cd5 |
| LibreLoop | c6e9cc1 (Trio's pin, from loopandlearn/LibreLoop; also the current tip of this repo's legacy-loopkit-compat). The latch code (notifySensorAttentionIfNeeded, LibreLoopSensorLifecycle, LibreLoopSensorMonitor) is unchanged on this repo's main @ 8c40f53, which every LibreLoop source link below points to. main also records CGM events when the flag latches (see Impact) |
| RoundWhiteDiscKit | e66da6c, the same revision this repo's main pins |
| Pump | Omnipod DASH |
Log and screenshots
📎 trio-libreloop-2026-10-30-redacted.zip (Trio Share logs export, 1.2 MB zipped, 54,815 lines, 30 Oct 00:00–19:03 AEDT)
Note
Redacted line-for-line, so every line number below matches the file. Removed: Nightscout host and server IPs, sensor serials (SENSOR-A = the flagged sensor, SENSOR-B = its replacement), pairing challenges, keys and PINs, receiver ID, Bluetooth addresses, and the app container ID. Glucose, insulin and carb data are kept.
| 15:57 Home | 15:57 CGM page | 16:09 Home |
|---|---|---|
![]() |
![]() |
![]() |
| 5.2 mmol/L, fresh, looping | Sensor failed, Bluetooth connected | 4.7 mmol/L, looping, still flagged |
To reproduce
Not deterministic: it needs the sensor to write an event-log entry with errorData = 8 while it is still active. In this log that happened during minute 8083 (≈13:22), right after a ~5 min app suspension, two notification-off writes from the silence watchdog, and the sensor dropping an authenticated connection (13:22:20).
An existing unit test on main already exercises this path: testReplacementRecordsTheSensorEndOnce feeds handlePatchStatus a frame with errorData = 8 and patchState = 4 (the helper's default, L182-L188) and asserts that the sensor end is recorded. sensorLifecycle is then .failed and stays there for repeated frames, whether or not actionable glucose keeps arriving. Suggested fix 1 would change that test's expectation.
Expected behavior
- A Sensor failed state needs corroboration that the sensor actually stopped:
patchStateleaving 4 (3/5/6/7/8), or glucose no longer arriving. - An
errorDatacode from an event-log entry whilepatchState == 4raises at most a soft, retractable Check sensor notice, ase0282d0already does for code 7. - The same event entry doesn't re-alert when it repeats or after an app relaunch.
Sequence
| Time (AEDT) | What happened | Log line |
|---|---|---|
| 13:16:51–13:21:45 | Phone suspended; nothing logged for 294 s | 38890–38891 |
| 13:21:45–13:22:05 | Silence watchdog sends the day's only two notification-off writes. The second comes from a monitor whose stream had already ended and lands on the new connection | 38891, 38895, 38975 |
| 13:22:01 | Patch status still carries the old entry: currentLC=8083 lifeCount=8 |
38971 |
| 13:22:20 | Sensor drops a live authenticated connection itself ("The specified device has disconnected from us"). Only time all day | 38980 |
| ≈13:22 | Sensor writes a new event-log entry at minute 8083 with errorData = 8 |
inferred |
| 13:24–13:34 | Connect → no-stream cycles; patch-status reads time out | 39064–39719 |
| 13:36:12 | First patch status after the event: currentLC=8097 lifeCount=8083 state=active → sensor attention: replaceSensor (rawAttention=replaceSensor, sensorError=terminated, patchState=active, replace=true); alert issued |
39952–39955 |
| 13:47–13:51 | 4 notDisplayable readings during a post-meal rise; no new event |
40619–40815 |
| 13:36–16:08 | 70 patch statuses, all lifeCount=8083 state=active; no further sensor attention lines |
39954–47615 |
| 16:08:48 | Last SENSOR-A reading (83 mg/dL) | 47626 |
| 16:09:30 → 16:10:37 | SENSOR-B scanned → paired; flag cleared | 47682, 47929–47930 |
| 16:18:32 | SENSOR-B's entry moves lifeCount 1 → 8 in the same packet its state goes raw(2) → active, then stays at 8 |
48205 |
Why it latches
1. Patch-status bytes 0–6 are the latest event-log entry.
Evidence
- SENSOR-A:
lifeCount=8in all 405 patch statuses from 00:01 to 13:22, thenlifeCount=8083in all 70 from 13:36 to 16:08, whilecurrentLifeCountkept advancing andstate=activethroughout. - SENSOR-B:
lifeCount=1whilestate=raw(2), changing to 8 exactly when the state went active (line 48205), then frozen. - RoundWhiteDiscKit's docs say the patch-status record shares the
lifeCount/errorData/eventData/indexshape of the sensor's event-log record:protocol.mdL426-L428. - The kit's captured field frame
9d280000e0000704e4381300decodes tolifeCount 10397withcurrentLifeCount 14564(bytes 8–9), an entry about 69 h old:DataPlaneTests.swiftL291-L301. - DiaBLE (
Libre3.swift,parsePatchStatus) labels these fields event life count, event error data, event data and last event index. Juggluco (Libre3GattCallback.java,libre3processpatchstatus) usesindexto request missing event-log entries.
2. RoundWhiteDiscKit treats that entry's errorData as a current condition. Libre3SensorAttention.init maps 8 → .replaceSensor regardless of patchState. The code describes the mapping as inferred (L46-L55), and neither DiaBLE nor Juggluco assigns meanings to errorData values.
3. LibreLoop's latch only re-evaluates on change. notifySensorAttentionIfNeeded returns early when attention == lastSensorAttention (L390) and persists sensorNeedsReplacement (L406-L414). LibreLoopSensorLifecycle.compute then returns .failed before rated wear. For the same sensor the only reset is a different attention value; otherwise only applyPairingOutcome clears it.
Was the sensor actually failing?
Not terminated, by every measure in the log: patchState 4 in all 70 packets after the event, 117 good readings after 13:36 with a smooth 5-minute history, last reading at 16:08:48, and no further event logged. The Bluetooth trouble was phone-side: SENSOR-B hit the same connect → no-stream pattern 16:30–17:13, about 20 minutes after pairing.
Note
The errorData = 8 entry is authenticated sensor data, written in the same minute as the sensor's only self-initiated disconnect of the day, so something happened inside the sensor. LibreLoop doesn't log the entry's eventData or index and never fetches the event log, so what it was can't be determined from this log.
Impact
- In Trio (where this was observed): display plus one time-sensitive alert; dosing was unaffected. Trio stores
hasValidSensorSessionbut never reads it (PluginSource.swiftL244-L245), never readsisInoperable, and LibreLoop kept forwarding glucose. The real cost was a working sensor removed ~9 days early, plus ~70 min of warm-up without closed loop. - In Loop: per
e0282d0's description, a latched replace marks the CGM inoperable and stops Loop. On Loopnext-dev,isInoperableputs a device issue on the loop status (Automation is unavailable), so a Loop user would be told to replace a sensor that is still delivering good readings. We haven't confirmed whether dosing actually stops:LoopDataManager.loop()doesn't readisInoperableorhasValidSensorSession. - On
mainspecifically:recordSensorStateIfChanged(from 04b0c03) also emits a.sensorIssueCgmEvent withisFatal: trueand a.sensorEndas soon as the flag latches. The working sensor's session would be recorded as ended at 13:36.
Suggested fix
- Don't latch on
errorDatawhilepatchState == 4. Downgrade to a retractable.checkSensor, ase0282d0does for code 7 (widening the.checkSensoralert case at L429 so it still raises a notice), and keep the hard latch for non-active patch states. A sensor that really dies stops sending glucose, which already shows Signal loss after 6 min (L36) and stops the host looping on stale data. - Persist the handled event identity (
lifeCount+index) so a repeated entry or an app relaunch doesn't re-alert.lastSensorAttentionis in-memory only (LibreLoopCGMManager.swiftL70). Optionally auto-clear once actionable readings newer than the event keep arriving. - (RoundWhiteDiscKit, optional) Expose the entry's age (
currentLifeCount − lifeCount) and don't derivesensorAttentionfromerrorDataalone whenpatchStateis active.
Related problems found along the way
-
discardSensorresetslastSensorAttentionbut notsensorNeedsReplacement/sensorEndedNormally, andapplyNFCResponsedoesn't clear them either. So the new sensor shows Sensor failed from NFC scan until the pairing outcome (16:09:30 → 16:10:37 in this log, lines 47682 and 47930). - The silence watchdog only checks cancellation at the top of its loop, so a finished monitor's in-flight tick sent a patch-status read (13:22:00) and a glucose notify-off (13:22:05, line 38975) onto the new connection, 15 s before the sensor dropped it. Possibly what provoked the event; not proven.
- The patch-status log line omits
errorData,eventData,indexand the disconnect-reason bytes. 04b0c03's.sensorIssuemetadata now recordserrorDataandpatchState.eventData,indexand the disconnect reasons only appear in the in-memory developer read inspector (L466-L475; last 300 reads, never persisted). They aren't logged or exported, so an entry like the one at minute 8083 can't be identified afterwards.
Additional context
- Precedent:
e0282d0(15 Jul) fixed the same false latch for code 7 and noted "Terminal replace (code 8, terminated) is unchanged." The only earlier field evidence for code 8 in this repo's history (16d74de, 1 Aug) came from a sensor shutting down after a clean end of wear. - Open questions: what does
errorData = 8mean inside an event-log entry? Has anyone else seenerrorData 8withpatchState 4on a Libre 3 / 3 Plus, and what happened next?
Generated by Claude Code



Activity
aidanlane commentedon Oct 30, 2026
@ps2
I initially reported it here: nightscout/Trio#1698
Don't fail a sensor over an event-log error while it is still active (L…