You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Hi, sometimes I need to take out air bubbles from the reservoir, after that the level of it is less than before, then i have a problem because the loop think that the insulin has been deliver and says "Eventually -1700mg/dl" I read that you can solve it going to the insulin delivery and delete the last half hour. I tried it, like when I put from error carbohydrates 2 times, I push the graph, then push edit and then appears "-" in red colour at the beginning of every line. the problem is that in the reservoir doesn’t appears the "-"sign, I push edit but nothing change....
Recent versions of Loop reinstate event deletion, and recognize prime events if you are using event history as your preferred data source.
How are you getting the bubbles out? By doing a prime?
On Jul 17, 2017, at 3:44 AM, davidmlb ***@***.***> wrote:
Hi, sometimes I need to take out air bubbles from the reservoir, after that the level of it is less than before, then i have a problem because the loop think that the insulin has been deliver and says "Eventually -1700mg/dl" I read that you can solve it going to the insulin delivery and delete the last half hour. I tried it, like when I put from error carbohydrates 2 times, I push the graph, then push edit and then appears "-" in red colour at the beginning of every line. the problem is that in the reservoir doesn’t appears the "-"sign, I push edit but nothing change....
I had selected event history.
Thanks
—
You are receiving this because you are subscribed to this thread.
Reply to this email directly, view it on GitHub, or mute the thread.
I getting out the bubbles, like when i change the reservoir, takein it out, an manally pressing to getting out, the same proces but not filling with insulin, so the total insulin will be less than before.
then is not posible to delete? and why when i did it loop consider the insulin hes been delivered?
Activity
ps2 commentedon Jul 17, 2017
davidmlb commentedon Jul 17, 2017
I getting out the bubbles, like when i change the reservoir, takein it out, an manally pressing to getting out, the same proces but not filling with insulin, so the total insulin will be less than before.
then is not posible to delete? and why when i did it loop consider the insulin hes been delivered?
thanks
ps2 commentedon Jul 17, 2017
You will need to move the pump motor somehow after you manually press on the reservoir. Are you using a prime or rewind+prime to do that?
If you are using Loop v1.3.2 or v1.3.3, then Loop will recognize prime events and not count them toward delivery.
davidmlb commentedon Jul 18, 2017
Yes, I do rewind like a reservoir change, i don't know what's prime? it's a fill the cannula?
the versión is the 1.3.3
davidmlb commentedon Aug 3, 2017
Whith new version 1.4 i can delete.
Solved
Merge pull request #522 from tidepool-org/ps2/LOOP-4211/therapy-setti…