I have maximum bolus in Loop set to 1U (and 6U on the pump itself) and was just able to do a 1.5U bolus from Loop.
Activity
CrushingT1D commentedon Jun 15, 2016
I tested this by setting my max to 0.1u in app and giving myself a 0.2u bolus and it gave the bolus as well.
loudnate commentedon Jun 16, 2016
Yep –
maxBolusis never used... 😊 https://github.com/loudnate/Loop/blob/master/Loop/Managers/DoseMath.swift#L131amazaheri commentedon Aug 21, 2016
Related to this would be great to validate carb entry, yesterday Sam by mistake keyed 5523 gr vs 55 and Loop suggested 553 units.
Fixing a crash when displaying the max bolus error
Improving VoiceOver accessibility for the HUD and Bolus screens (#192)
Merge pull request LoopKit#33 from loudnate/swift2.2
Fix copy-frameworks script; missing quotes around environment variable (
Fake BG sweep (LoopKit#33/R29): exercise the high-temp range Jeremy's…