You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
{{ message }}
Repository navigation
Bug: Incorrect pod suspended message when bolusing after comms error #2198
Updated Status
Leaving the issue open until the submodule updates associated with this issue are added to LoopWorkspace for released and development code
Describe the bug
Using my regular Looping phone, I tried to bolus and Loop reported Pod was suspended (it was not).
Repeated the bolus attempt and it went through successfully.
I have seen comments about this intermittent error on Facebook, but never got a Loop Report.
the Device Communication Log showed there was IO error (“messageIOException("Could not write $msgType: unsentWithError(OmniBLE.PeripheralManagerError.timeout([...”)
this was on the last get status with a response before the bolus was attempted
this combination resulted in the incorrect warning about the pod being suspended on the subsequent bolus attempt
To Reproduce
This is an intermittent error.
Expected behavior
Do not show user a message that pod is suspended when it is not.
Screenshots
n/a
Phone
iPhone 15 pro, iOS 17.5.1
Loop Version
essentially the same as Loop 3.4.1 (before the release)
CGM
Dexcom G7 using Dexcom app
Pump
Omnipod DASH
Additional context
I was super busy at the time. I sent the Loop report to Joe (itsmojo).
This issue should be fixed with the two updates to OmniBLE and OmniKit.
Leaving the issue open until these submodule updates are added to LoopWorkspace for released and development code.
This is similar or identical to a similar issue I was going to file today, with non-Dash Omnipods. New pump, first bolus, inexplicably suspended immediately while not being suspended, and acts as though it is suspended (any subsequent dosing attempts fail until I suspend and resume the pod)
Activity
marionbarker commentedon Jul 27, 2024
This issue should be fixed with the two updates to OmniBLE and OmniKit.
Leaving the issue open until these submodule updates are added to LoopWorkspace for released and development code.
caitp commentedon Sep 22, 2024
This is similar or identical to a similar issue I was going to file today, with non-Dash Omnipods. New pump, first bolus, inexplicably suspended immediately while not being suspended, and acts as though it is suspended (any subsequent dosing attempts fail until I suspend and resume the pod)
marionbarker commentedon Mar 14, 2025
Updated for main in V3.4.4 release. It has also been added to Trio and iAPS.