Describe the bug
I'm on steroids so I have increased my basal rate across the board. I updated my basal rates through the Loop interface and they appear to be written to the pump;however, there is a mismatch between the Insulin Delivery display and the actual current basal. Currently, my basal is set at 1.4 U/hr (12 AM-12PM) but the display reads as Scheduled Basal 0.7 U/hr. This is interesting because that is my typical basal rate without steroids. It is also the last basal rate that I likely entered manually into the pump and not via Loop's interface, although guessing a little here.
Attach an Issue Report
I'm happy to email the Issue Report--my github app keeps crashing To Reproduce
Steps to reproduce the behavior:
Go to Settings > Minimed 523 > scheduled basal
Check basal rates under Therapy Settings--mismatch here
Check 523 for basal rate schedule written to pump
Press Esc on pump and check current basal rate
Expected behavior
I'd expect the Scheduled Basal display to match what is in Therapy Settings and what shows on the pump
Conclusion: Display on MDT Pump screen shows the value initially exchanged with pump and is not updated when Therapy Settings are changed in Loop.
Next test (same dev configuration as above).
Test Details:
delete MDT pump
add MDT pump (still with settings of 1.25 U/hr on pump itself and in Loop Therapy settings)
wait for green loop
Loop MDT screen now shows the true 1.25 U/hr
Repeat the modification to therapy setting - set to 1.15 U/hr in Loop
Confirm pump reports 1.15 U/hr
Confirm Loop MDT screen continues to show original schedule (1.25 U/hr)
Also confirmed that the Loop Report indicates scheduled basal equivalent to pump and to therapy setting.
So it is a display problem with the MDT display.
Activity
marionbarker commentedon Feb 26, 2024
I confirm this behavior with Loop 3.2.3 and a MDT 515 pump.
I will test with dev in a separate comment.
Configuration:
Test Details:
marionbarker commentedon Feb 26, 2024
I confirm that the display is incorrect with dev as well.
Configuration:
Test Details:
marionbarker commentedon Feb 26, 2024
Conclusion: Display on MDT Pump screen shows the value initially exchanged with pump and is not updated when Therapy Settings are changed in Loop.
Next test (same dev configuration as above).
Test Details:
Also confirmed that the Loop Report indicates scheduled basal equivalent to pump and to therapy setting.
So it is a display problem with the MDT display.
github-actions commentedon Mar 28, 2024
This issue is stale because it has been open for 30 days with no activity.
marionbarker commentedon Mar 28, 2024
Bump
marionbarker commentedon Jul 19, 2024
This was fixed with the release of Loop 3.4.
This issue can be closed now.